How it works
External-standard assay compares the sample response with a reference standard of known purity prepared at a similar concentration. Both concentrations are expressed as the final concentration injected, after every dilution step, so dilution factors must be applied consistently.
Use the standard purity on the same basis as the result: as-is purity for an as-is result, or correct for water or loss on drying separately. For salts, include the free-base correction in the standard purity (purity × MW free base / MW salt) when the label claim is expressed as the free base.